(mV) (mV)
4.1.4 Depress TI – mV pushbutton –1100 –1098 to –1102
–1300 –1298 to –1302
4.2 pH Indication Tests
4.2.2 TI Controls(s):
Depress STBY, pH, and EXP
pushbuttons
Set TEMPERATURE control to
30 adjust STANDARDIZE
CONTROL for TI reading of 7.00 ck( ) NA
4.2.3 (DC millivolt source)
(mV) (pH) (pH)
300.8 2.00 1.99 to 2.01
180.5 4.00 3.99 to 4.01
60.15 6.00 5.99 to 6.01
–60.15 8.00 7.99 to 8.01
–180.5 10.00 9.99 to 10.01
–300.8 12.00 11.99 to 12.01
4.3 pH Calibration Tests
4.3.4 TI control(s):
Depress pH, EXP, and STBY
pushbuttons
TI TEMPERATURE
and SLOPE controls to OFF
Set MIDSCALE VALUE control to 7
Ensure Zero pushbutton on
rear panel is released
Release STBY pushbutton
Adjust needle to 0 on the +
0.7 scale 0 ck( ) NA
Depress STBY pushbutton
TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO. MFG. MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20SC–42
Page 2 of 3
Beckman Instruments 2000 Ion Analyzer
NAVAIR 17–20SC–42